Shop Temperature Instruments
Sensors and probes
Everything else on this page depends on the sensor. Choosing it is a trade-off between temperature range, accuracy, response speed and how much punishment the probe has to take.
Thermocouples
Suitable for: Liquids · Gases · Steam (in a pocket) · Solids (surface probes)
A thermocouple is two wires of different alloys joined at the tip. Any temperature difference between that tip and the cold end of the wires generates a small voltage, and the instrument reads the voltage and converts it back to a temperature. Because it has to know the temperature at the cold end to do that, every thermocouple instrument measures its own terminal temperature and corrects for it, which is called cold junction compensation.
Type K is the general-purpose workhorse and covers most industrial work. Type T suits low temperatures and food and refrigeration use, Type J is common on older plant and plastics machinery, and Type N holds up better than Type K at sustained high temperatures. Thermocouples are cheap, tough, fast and cover an enormous range, but they are less accurate than a Pt100 and the extension wire has to be the matching type, with the polarity the right way round.
Pt100 and Pt1000 sensors (RTDs)
Suitable for: Liquids · Gases · Steam (in a pocket) · Solids (surface probes)
A resistance temperature detector measures temperature by the change in a metal’s electrical resistance. Platinum sensors are the standard: a Pt100 reads 100 Ω at 0 °C and a Pt1000 reads 1000 Ω, and the resistance climbs in an almost straight line as it warms. That near-linear behaviour is what makes RTDs the accurate, stable, repeatable choice for process and laboratory measurement up to around 600 °C.
Wiring matters more than most people expect. A two-wire connection adds the resistance of the cable to the reading, which can be worth a degree or more on a long run. Three-wire cancels most of that error and is the usual industrial choice; four-wire cancels it completely and is used for reference work.
Thermistors
Suitable for: Liquids · Gases · Solids (surface probes)
A thermistor is a ceramic semiconductor whose resistance changes steeply, and not at all linearly, with temperature. That steepness gives excellent sensitivity across a narrow band, which is why thermistors are used in refrigeration, HVAC, appliances, incubators and equipment protection rather than in wide-range process work.
Probe styles, pockets and accessories
The sensing element is only part of the order. We supply handheld probes in Type K and Pt100, surface probes for checking a plate or a bearing housing, needle and penetration probes for food, fixed probes with a connection head for permanent installation, and specialist probes for the plastics industry. Thermowells and pockets let you withdraw the probe without breaking into the process. We also stock thermocouple extension wire, compensating cable, plugs, sockets and panel connectors in the matching alloys.
Custom RTD and thermocouple probes are made in New Zealand to your specification by Teltherm Instruments Ltd. A non-standard length, sheath material, process connection or termination is a build rather than a special import, which usually means it arrives sooner and can be repeated exactly next time.


Thermometers, indicators and loggers
These are the instruments that show you the temperature, whether you are standing in front of the process or reading a month of data at your desk.
Analogue dial thermometers
Suitable for: Liquids · Gases · Steam (in a pocket)
A bimetal dial thermometer uses two metals bonded together and wound into a coil. The metals expand at different rates, so the coil winds or unwinds as the temperature changes and turns the pointer. It needs no power, survives a plant room, and shows you a reading when everything else is off. Gas- and liquid-filled versions use the expansion of a sealed fluid instead, and a capillary can put the dial some distance from the point being measured.
Our bimetal dial thermometers are made in New Zealand by Teltherm Instruments Ltd, the only New Zealand manufacturer of bimetal analogue temperature gauges, so dial sizes, stem lengths, ranges and custom scales can be made to order.
Fit a dial thermometer into a pocket rather than directly into the process wherever you can. It can then be removed or replaced without draining the line.
Digital thermometers
Suitable for: Liquids · Gases · Solids (surface probes)
Handheld digital thermometers accept interchangeable probes, so one instrument covers immersion, surface, penetration and air measurement. Look at the input types it accepts, the resolution, whether it holds minimum, maximum and average, and whether it takes two inputs at once, which lets you read a difference directly rather than working it out.
Infrared thermometers
Suitable for: Liquids (surface only) · Solids (surfaces)
An infrared thermometer reads the heat radiated from a surface without touching it, which makes it the only practical option for something moving, live, very hot or out of reach. Two settings decide whether the reading is right. Emissivity describes how efficiently the surface radiates; shiny bare metal radiates poorly and will read far too low unless the setting is corrected or the spot is painted or taped. Distance-to-spot ratio tells you how big an area the instrument is averaging, so at a distance you may be measuring the wall behind the target as well.
Temperature and humidity loggers
Suitable for: Liquids · Gases · Solids (surface probes)
A logger records readings at a set interval and stores them, so you get the pattern rather than a single number. That is what proves a cool store held temperature overnight, shows the swing in a drying oven, or documents conditions in a vehicle or shipment. Options range from single-use temperature tags to multi-channel units for temperature, humidity, pressure and process signals, with USB, wireless or networked download.
Transmitters, switches and controllers
A sensor on its own produces a fragile, low-level signal. These instruments turn it into something your system can use, or act on it directly.
Temperature transmitters
Suitable for: Liquids · Gases · Steam (in a pocket) · Solids
A temperature transmitter sits at the sensor, usually in the connection head, and converts its millivolt or resistance signal into a robust 4–20 mA or digital output. Doing that conversion at the sensor rather than in the panel matters: the weak signal travels centimetres instead of tens of metres, so electrical noise, cable resistance and junction effects stop being a problem. Most transmitters are configurable, so the same unit handles several sensor types and ranges.
Temperature switches
Suitable for: Liquids · Gases · Solids
A temperature switch changes the state of a contact at a set temperature, to raise an alarm, start a fan, or shut down equipment before it is damaged. Mechanical versions use a filled bulb and capillary acting on a switch mechanism; electronic versions use a sensor with set points entered on a display.
Temperature controllers
Suitable for: Liquids · Gases · Solids
A controller closes the loop. It reads the sensor, compares it to the set point and drives the output to hold the process there. Simple on/off control cycles the heat above and below the set point, which is fine where a little swing does no harm. PID control adjusts the output continuously from the size of the error, how long it has persisted and how fast it is changing, which holds a much tighter set point and is what you need on extruders, ovens, furnaces and jacketed vessels. Most modern controllers can tune their own PID terms on request.


Thermostats and heating control
Alongside process instruments, we supply the controls that heat buildings: thermostats, radiator valves, actuators and heating elements for hydronic, underfloor and electric systems.
Room and underfloor thermostats
Suitable for: Gases (air) · Solids (floor sensor)
Mechanical thermostats switch on a bimetal element and are simple, cheap and reliable. Digital thermostats add an accurate sensor, a programmable schedule and, on Wi-Fi and touchscreen models, remote control and energy reporting. Underfloor models take a floor sensor as well as the room sensor, so the floor temperature can be limited to protect the covering while the room is controlled on air temperature.
Hydronic controls, radiator valves and actuators
Suitable for: Liquids (heating water)
Hydronic systems move heat as hot water. Thermostatic radiator valves control each room individually by modulating the flow into that radiator, and motorised actuators on manifold valves let a central thermostat or controller open and close individual underfloor circuits. Zoning a system this way is usually the cheapest useful improvement in both comfort and running cost.
Heating elements
Suitable for: Liquids · Gases · Solids
We supply cartridge, band, strip and immersion heating elements for process equipment and tanks. The figure that decides element life is watt density, the power per unit of surface area: too high for the medium and the element scales, coats or burns out early.
Humidity measurement
Humidity affects product quality, comfort, condensation and corrosion, and it is almost always measured alongside temperature, because the two are inseparable.
Which humidity measurement do you need?
Suitable for: Gases (air and process gases)
- Relative humidity (%RH) is how much water vapour the air holds as a percentage of the most it could hold at that temperature. It is the everyday measurement, and it changes as the air warms or cools even when no moisture is added.
- Dew point is the temperature at which that vapour would start to condense. It does not move when the air warms or cools, which is why it is used for compressed air, drying and anywhere condensation is the risk.
- Absolute humidity is the mass of water vapour in a given volume of air, such as grams per cubic metre.
Humidity instruments
Suitable for: Gases (air and process gases)
Nearly all modern humidity instruments use a capacitive polymer sensor: a thin film absorbs water vapour and its capacitance changes in proportion to relative humidity. We supply wall and duct-mounted transmitters, handheld meters and combined temperature and humidity loggers. Humidity sensors are affected by dust, solvents and ageing more than most, so they need regular calibration.

Choosing a temperature sensor
Start with the range and the accuracy you actually need, then let the process narrow the rest.
What we will ask you
- the temperature range, and the accuracy the process or audit requires
- the medium, and whether the probe can touch it
- the process connection, immersion depth and whether a pocket is needed
- how fast the reading has to respond
- what the reading feeds: a local display, a logger, a control system or a controller
- hazardous area, hygienic and certification requirements.
Which sensor suits the job?
| Sensor | Liquids | Gases | Steam | Solids (surfaces) | Best for |
|---|---|---|---|---|---|
| Thermocouple | ✓ | ✓ | ✓ (in a pocket) | ✓ | Wide range, high temperatures |
| Pt100 / Pt1000 | ✓ | ✓ | ✓ (in a pocket) | ✓ | Accuracy and stability |
| Thermistor | ✓ | ✓ | – | ✓ | Sensitive over a narrow band |
| Infrared | Surface only | – | – | ✓ Surface only | No contact with the target |
| Bimetal dial | ✓ | ✓ | ✓ (in a pocket) | – | Local reading, no power |
| Filled system | ✓ | ✓ | ✓ (in a pocket) | – | Dial away from the process |
| Humidity sensor | – | ✓ (air) | – | – | Relative humidity and dew point |
Calibration and repair
Our temperature, infrared and humidity calibration is accredited by IANZ to ISO/IEC 17025. We repair, adjust and certify dial thermometers, thermocouples, Pt100 and Pt1000 probes, surface probes, digital thermometers and infrared thermometers, with on-site calibration available for systems that cannot come to the laboratory. Liquid-in-glass thermometers are the one exception we do not certify. See Calibration Lab for more.
Frequently asked questions
Are your thermometers made in New Zealand?
Our bimetal dial thermometers are, and so are our custom RTD and thermocouple probes. Both are manufactured by Teltherm Instruments Ltd, the only New Zealand manufacturer of bimetal analogue temperature gauges. Non-standard lengths, ranges, connections and scales are made to order.
Thermocouple or Pt100?
Pt100 for accuracy and stability up to around 600 °C, and where a repeatable reading matters for quality or audit. Thermocouple for higher temperatures, a faster response, a tighter space or a tougher installation, and where a degree either way is acceptable.
My readings changed when I extended the cable. Why?
On a thermocouple, ordinary copper cable creates a second junction at the joint and shifts the reading; you need matching extension or compensating cable, connected the right way round. On a two-wire Pt100, extra cable adds resistance that the instrument reads as extra temperature; change to a three-wire connection.
Why does my infrared thermometer read low on stainless steel?
Because shiny metal radiates heat poorly. Set the instrument’s emissivity to match the surface, or stick a piece of matt tape on the target and measure that instead once it has reached the same temperature.
What is a thermowell for?
It is a closed tube fitted into the process that the probe drops into. You can remove or replace the probe without draining or shutting down the line, and the process never touches the sensor. It does slow the response a little, which is normally worth accepting.
Do I need a logger or a transmitter?
A logger records readings for you to review later, which is what you want for proof and investigation. A transmitter sends a live signal to a control or monitoring system, which is what you want for control and alarms. Plenty of sites use both.
How often should temperature instruments be calibrated?
Working instruments every 12 months, with an ice point check every three to six months as a quick confidence check in between. Reference thermometers can go up to five years if ice point checks are done every six to twelve months. Humidity and infrared instruments every 12 months.