Book a Calibration
What we calibrate
Our accredited scope covers four measurement areas, and we handle gas monitors and backflow testers alongside them. The exact ranges and uncertainties we are accredited for are listed in our IANZ Scope of Accreditation.
Pressure
We repair, adjust and certify analogue and digital gauges, transducers and transmitters, and differential and low-pressure instruments, from −1 bar vacuum to 700 bar. Your instrument is compared with a reference standard at several points across its range, and readings are taken both as the pressure rises and as it falls, so the certificate shows hysteresis as well as accuracy. The test medium depends on the range and the instrument: air, nitrogen, oil or water. Oxygen-clean service is available on request.
Temperature
We repair, adjust and certify dial thermometers, thermocouples, Pt100 and Pt1000 probes, surface probes and digital thermometers. Calibration is by comparison: your sensor and a reference thermometer are held together at a stable temperature in a stirred liquid bath or a dry-block calibrator, and their readings are compared at each test point. Liquid-in-glass thermometers are the one type we do not certify.
Infrared thermometers
An infrared thermometer is calibrated against a blackbody source: a cavity held at a known temperature with a known, near-perfect emissivity. Because the reading also depends on the instrument’s own emissivity setting and its distance from the target, the certificate records the conditions the calibration was carried out under.
Humidity
Humidity instruments are calibrated in a controlled chamber at set humidity points and compared against a reference hygrometer. Humidity sensors are more exposed than most: dust, solvent vapour and simple ageing all shift the reading, which is why they need checking on a regular interval rather than only when something looks wrong.
Gas monitors
Homershams, in association with Teltherm Instruments Ltd, offers service, repair, calibration and IANZ certification of gas detectors. The detector is zeroed in clean air, exposed to certified gas of a known concentration, adjusted to read it correctly, and its alarm points are checked and recorded.
Backflow testers
Backflow prevention devices protect the drinking water supply, and the test kits used to verify them have to be calibrated themselves for the results to stand. We calibrate backflow test kits and supply replacements. See Backflow Testers for details.


In the laboratory or on your site
Some instruments can come to us and some cannot. Both are covered, and the choice usually makes itself.
Laboratory calibration
Send the instrument to us wherever you can. In the laboratory the full set of reference standards is on hand, conditions are controlled, and the spare parts are in the building, so an instrument that fails can often be adjusted or repaired and certified in the same visit rather than coming back to you with a fail and a quote.
On-site calibration
Integrated systems, large installations and instruments in a process that cannot be interrupted are calibrated where they are. Our field team brings the reference equipment to you, works around your shutdown or maintenance window, and issues the same certificates. On-site work is arranged in advance so the right standards come with the technician.
Service, repair and replacement
Calibration tells you whether an instrument is within specification. It does not, on its own, get you working again. Because we supply the instruments as well as certify them, we can repair, adjust or replace most of what comes through the laboratory, which is usually faster than sourcing it elsewhere. See Service & Repair for more.
Accreditation and traceability
Two things make a calibration certificate worth anything: an unbroken chain back to a national standard, and an independent assessment that the laboratory is competent to do the work.
What traceability means
Traceability means your result can be linked back to a national or international standard through a documented, unbroken chain of calibrations, with an uncertainty stated at every step. Each link in that chain adds a little to the total uncertainty, which is why the chain is kept as short as it sensibly can be. Without traceability, a certificate says only that two instruments agreed with each other, with no way of knowing whether either was right.
The chain above us
At the top sits the International System of Units, maintained through the International Bureau of Weights and Measures (BIPM) under the Metre Convention, which runs the international comparisons that keep countries in step with one another. Beneath that are the national metrology institutes; in New Zealand that is the Measurement Standards Laboratory. We calibrate our reference standards against those national standards, and then calibrate your working instruments against our references.
IANZ accreditation to ISO/IEC 17025
ISO/IEC 17025 is the international standard for testing and calibration laboratories. It covers both the technical competence to produce a valid result and the management system that keeps producing it. International Accreditation New Zealand (IANZ) assesses us against that standard, and our accreditation covers pressure, temperature, infrared and humidity. In 1980 we became the first pressure laboratory in New Zealand to be certified by Telarc for pressure calibration.
Recognised internationally
IANZ is a signatory to the International Laboratory Accreditation Cooperation (ILAC) Mutual Recognition Arrangement. Our IANZ laboratory has mutual recognition with accreditation bodies in over 65 countries, which means a certificate issued here is accepted by regulators and customers in those markets without the work being repeated.
Our calibration capabilities
The ranges and best measurement uncertainties we are accredited for are published in our IANZ Scope of Accreditation. If you are not sure which calibration you need, or what a figure on a certificate means, our team will go through it with you.

How often to calibrate
Your quality system or the instrument documentation should set the interval. Where neither does, the table below is a sensible starting point. Two rules override any schedule: calibrate after an event that could have damaged or shifted the instrument, and calibrate whenever you have stopped trusting the reading.
Recommended intervals
| Instrument | Full calibration | In between |
|---|---|---|
| Pressure gauges and transmitters | Every 12 months | – |
| Working temperature instruments | Every 12 months | Ice point check every 3–6 months |
| Reference temperature instruments | Every 5 years | Ice point check every 6–12 months |
| Humidity instruments | Every 12 months | – |
| Infrared thermometers | Every 12 months | – |
| Gas detectors | Every 6 months | Bump test before each use |
| Backflow test kits | Per your testing authority | – |
An ice point check is a quick reading in a properly made ice bath at 0 °C. It takes minutes, needs no equipment you do not already have, and will catch a reference thermometer that has shifted long before its next full calibration is due.
We keep a reminder system that tells you when each of your instruments is next due, so a calibration schedule does not depend on somebody remembering.
Frequently asked questions
What is the difference between traceable and accredited calibration?
Traceable means the result links back to a national standard through an unbroken chain of calibrations. Accredited means an independent body, IANZ in New Zealand, has assessed the laboratory and confirmed it is competent to do that work to ISO/IEC 17025. An accredited calibration is traceable, and the way the uncertainty was calculated has been independently checked.
How long will my instrument be away?
It depends on the instrument and the workload at the time, and on whether a repair or adjustment is needed. Tell us when you need it back when you book it in and we will tell you what is realistic.
What happens if my instrument fails?
We will contact you before doing anything chargeable. In most cases we can adjust or repair the instrument and certify it, and if it cannot be brought back into specification we will tell you what a replacement would cost.
Can you calibrate instruments you did not supply?
Yes. We calibrate instruments regardless of where they came from, within the ranges published in our IANZ scope of accreditation.
Do you calibrate liquid-in-glass thermometers?
No. We calibrate and certify most other thermometer types, including dial thermometers, thermocouples, Pt100 and Pt1000 probes, surface probes, digital thermometers and infrared thermometers.
Is a bump test the same as calibrating a gas detector?
No. A bump test applies a small amount of gas to confirm the detector responds and alarms, and should be done before each use. Calibration uses certified gas of a known concentration to adjust the readings, and produces a certificate.